Z-STERILIZER

Surgical Instrument Disinfectant- Ideal for the sterilization of flexible endoscopes, ultra-sound probes and other reusable, non-invasive, heat sensitive medical devices.

Components Sodium Perborate+Activator organic acid
Corrosion inhibitors and stabilizers

Efficacy: Z-Sterilizer has independently proven against Viruses (including HIV and Hepatitis B&C), Bacteria (Ps.Aeruginosa), Mycobacterium (MIA, TB) Fungi and spors. These tools show rapid Sterilant properties across a wide spectrum of Micro-organisms marking Z-Sterilizer ideal for Endoscopy.

Compatibility:
Z-Sterilizer is compatible with flexible and stainless steel rigid Endoscopes. Z-Sterilizer should only be used in AER's with sainless steel or inert material valves and pipe work. Routine enzymatic pre-cleaning of instruments to remove fixed proteins and other biofilms is unnecessary. Pre-cleaning with non-enzymatic detergent saves cost.

Area of Application:
- Add 16.2g of Z-Sterilizer powder per 1 liter of Luke warm water (810g/50 liter) stir until the powder fully dissolved and solution is Blue/Green color
- Z-Sterilizer must be used at 20'c-25'c for rapid activation (typically 10 minutes) use Luke warm water is defined as 30'c-35'c activated solution is ready to use
- Z-Sterilizer can be used in static bath(Tray) immerse instruments for 10 minutes after which they must be rinsed thoroughly in sterile water, dried and reused or stored as appropriate
- Z-Sterilizer solution may be discarded without special pre cautions, in the normal liquid waste disposal system after 24 hrs in accordenance with local authority

Mode of Action: Z-Sterilizer mode of action by oxidation of proteins by Hydrogen Peroxide and Peracetic acid.

Material Compatibility: BN-OPA has excellent material compatibility with glass, ceramic, stainless steel, non-ferrous metals, aluminum, rubber, latex, and silicone, synthetic materials such as polycarbonate and acrylic +glass

Storage and shelf life:
Store only in cool, humidity controlled conditions. Shelf-life 2 years from date of manufacture. Container must be tightly re-sealed to preserve integrity of powder Life of Solution
- 24 hours from activation, or
- Maximum 20 cycles in Automated Endoscope Reprocessor, or
- Until indicated by test-strip,
- As per good Infection Control practice solutions must be discarded after 24 hours use